Failure Analysis
Adopting low-voltage STEM and automated sample prep to perform IC failure
analysis
Bryan Tracy and Kirstin Alberi, Spansion; and Shams Tabrez, SELA
USA
Shrinking semiconductor device sizes have pushed many routine measurement
and analysis tasks beyond the resolution capability of scanning electron
microscopy (SEM). Transmission electron microscopy (TEM) can provide higher
resolution than SEM, but it is more expensive and more difficult to perform.
Scanning transmission electron microscopy (STEM) is finding growing acceptance
in semiconductor failure analysis laboratories as a viable alternative,
providing image resolution approaching that of a TEM but with a cost and
ease of use more like that of a SEM. However, analysts are reluctant to
take on the burden of preparing the types of thin samples associated with
STEM analysis. Automated sample preparation techniques developed for SEM
and TEM address this concern.
This article compares the ability of both TEM and STEM to analyze typical
semiconductor samples and describes an approach to performing automated
sample preparation that makes the process easy, fast, and reliable.
SEM/TEM/STEM Alphabet Soup
A SEM such as that pictured in Figure
1a forms an image by scanning a finely focused beam of electrons over
the sample surface and mapping, point by point, the intensity of various
signals emitted by the sample as a result of interactions between the
beam electrons and the sample atoms. If the spot formed by the beam on
the sample surface drops below a certain size, SEM resolution is limited
by the size of the region from which the mapped signal emanates. That
region is any position where beam
electrons scatter when they enter the sample. The size of this interaction
volume depends on many imaging conditions, foremost among which are the
particular signal chosen and the energy of the incident electron beam
(measured in kV). Most SEMs operate with beam energies of 20 to 30 kV.
At those energies, the interaction volume for some signal types can be
a cubic micrometer or more. While the use of low voltage when operating
with beam energies on the order of 1 kV reduces the size of the interaction
volume, low-energy electrons are more difficult to focus into a small
beam, limiting resolution.
A TEM such as that shown in Figure
1b is analogous to a film projector as it projects a magnified film
image onto a screen. A broad beam of electrons floods the imaged region
of a thin sample, while magnetic lenses form a real image from transmitted
electrons and project the image onto a fluorescent screen or other viewing
device. Image contrast can derive from a number of different interactions
between beam electrons and sample atoms, making image interpretation complex.
Since the TEM sample is very thin—typically a tenth of a micron
or less—the beam electrons have little opportunity to spread. TEM
resolution is limited primarily by the performance of the lenses in the
electron optical system. The higher beam energies of TEMs compared with
SEMs—in the 200- to 300-kV range—yield better resolution.
The STEM (shown in Figure 1c)
developed as a hybrid technique performed on modified TEMs or SEMs. As
in a SEM, the beam focuses on a small spot that scans over the sample.
The image is formed by mapping some signal intensity synchronously with
the scan. As in TEM, image information is extracted from electrons that
are transmitted through a thin sample. STEM has many of the advantages
of SEM: TV rate imaging, very-low-magnification (wide-field) imaging,
and easy electronic image rotation. However, it outperforms SEM in that
it offers stronger materials contrast, longer sample lifetime in the beam,
and lower effective contamination rates. Moreover, it does not require
charging or special handling of conductive coatings. Because STEM samples
are thin, beam spreading in the sample is minimal, and resolution is determined
primarily by the spot size formed by the beam on the sample. While STEM’s
resolution can approach that of a TEM, in practice some of its resolution
is often traded for increased beam current and signal strength.
The term “low kV” as applied to STEM can be a little confusing.
Low-voltage STEM as used here refers to the 20–30-kV regime, i.e.,
low relative to typical TEM beam energies. It is near the lower limit
of energies that will provide sufficient transmission through the sample.
STEM can also be performed at full TEM beam energies.
STEM can be performed on a dedicated instrument specifically designed
to perform the technique, but more often it is added onto a SEM or TEM.
In addition to being able to focus the beam on a spot and scan it over
the sample, it must have an appropriate detector on the transmission side
of the sample. STEM is becoming increasingly appealing because it can
be adapted to a SEM platform and achieves high-resolution imaging comparable
to that of a TEM. Since a SEM is designed to scan a focused beam over
a sample, the only modification required of it to function as a STEM is
the introduction of a transmitted-electron detector. Several such devices
are commercially available.
The electrons transmitted at any point that are illuminated by the scanning
beam may be divided into three regimes based on their interactions with
the specimen. These regimes can be visualized as three coaxial cones of
increasing angularity whose common vertices are at the point of transmission,
as illustrated in Figure 2.
 |
| Figure
2: Diagram of the bright-field, annular dark-field, and high-angle
annular dark-field functions of a STEM. |
The central cone contains electrons that have passed largely unaffected
through the sample. The intensity of the signal generated here is a function
of the mass thickness of the sample. The thicker or more massive the sample,
the more electrons are absorbed or scattered out of the cone and the weaker
the signal becomes. When using this signal, which is called the bright-field
signal, more-massive features are darker.
The intermediate cone contains electrons that have been scattered by
interactions with electrons of the sample atoms, and, if the sample is
crystalline, electrons that have been diffracted by its periodic structure.
(In a conventional TEM, the pattern formed by the diffracted electrons
is an important source of information about crystalline structure.) TEM
images formed by these electrons are called dark-field images and are
roughly the complement of the corresponding bright-field image. In other
words, most electrons that are diffracted or scattered out of the bright-field
signal end up in the dark field signal.
The third and widest cone contains electrons that have been scattered
through high angles by interactions with the atomic nuclei of the sample
(the same process that creates backscattered electrons in a SEM). The
intensity of this signal is primarily a function of the atomic number
of the sample. Images formed from it are known as Z-contrast images, and
STEM detectors designed to capture this signal are called high-angle annular
dark-field (HAADF) detectors.
Sample Preparation
Perhaps the most important reason that SEM has been the preferred technique
in failure analysis labs is that it requires only minimal sample preparation.
Especially since the appearance of variable-pressure and environmental
SEMs, a sample probably can be imaged if it fits in the chamber. TEM and
STEM, on the other hand, require small, extremely thin samples that can
be very difficult to create. Whole careers have been built around TEM
sample-preparation techniques. Failure analysts, for whom time is money,
have been reluctant to take on the burden of using a TEM or STEM if a
SEM could do the job.
In addition to the requirement that it be thin enough to transmit electrons,
the typical failure analysis sample must be extracted from a particular
wafer location (identified by an inspection tool) with sufficient accuracy
to ensure that it includes the targeted feature or defect. Conventional
preparation techniques consist of manually cleaving the wafer to a tiny
sliver containing the target and then painstakingly polishing the sliver
to remove material from both sides to reveal the target. This process
of sizing and polishing the wafer is typically performed manually. However,
manual processing can contaminate, damage, or even destroy a sample entirely.
 |
Figure
3: Schematic diagrams of the MC600 automated sample-preparation process:
(a) a coarsely cleaved sample,
(b) microcleaving of coarse sample, (c) automatic scribing of the
wafer edge, (d) shock wave induced from the opposite edge of the wafer
piece before cleaving, and (e) final cross-sectioning of the target. |
An automated sample-preparation process, presented schematically in Figure
3, avoids the risks associated with the manual procedure. First, a coarsely
cleaved sample (Figure 3a) is microcleaved (Figure 3b) using an MC600
cross-sectioning system from SELA (Upper Yokneam, Israel; and Sunnyvale,
CA). The purpose of the cleaving step is to achieve a wafer piece measuring
several millimeters per side with the target feature located approximately
10 µm from one edge. Then the system automatically navigates to
wafer defect coordinates provided by an inspection tool, after which the
operator uses an optical microscope to refine the position of the desired
cleave. Available software enables the operator to plan the cleaving process
and yield multiple samples from a single wafer. Once the target has been
pinpointed, the system automatically proceeds to scribe the wafer edge
(Figure 3c) and initiates each cleave with a carefully controlled shock
wave induced from the opposite edge of the wafer piece (Figure 3d). Finally,
final cross-sectioning of the target is performed (Figure 3e). The MC600
is pictured in Figure 4.
 |
| Figure
4: The MC600 cross-sectioning tool. |
After microcleaving, the sample is thinned in a SELA TEMpro tool, where
it is automatically glued to a specially designed stub. As illustrated
schematically in Figure 5a,
the system uses a precision diamond saw to remove the bulk of the wafer
piece and the unneeded portions of the stub. Glued to a TEM grid and gripped
in a clamp (Figures 5b-h), the final sample is approximately 25 µm
thick in the region of the target and ready for final processing. Although
several techniques have been developed that substitute various forms of
ion milling for polishing in the final stages of sample preparation, many
semiconductor failure analysis labs use focused ion beam (FIB) systems
to perform a final “polish.” FIB uses a finely focused beam
of relatively heavy ions, much like a tiny sandblaster, to remove material
with great accuracy and precision. However, although FIB systems offer
very precise control over material removal, they are inherently slow and
expensive.
Automated sample preparation is faster, easier, and more reliable than
its manual equivalent. The process can prepare a FIB-ready sample in less
than an hour with 90% reliability. Automated systems can be operated by
technician-level personnel, freeing them for other tasks during much of
the process.
Examples
STEM, TEM, and SEM micrographs of a transistor are presented in Figures
6a–6c, respectively. The STEM image (at 30 kV) and the TEM image
(at 200 kV) are of comparable quality. The individual components of the
transistor can be distinguished in both, and the contrast differential
between oxide and nitride is roughly equal. The resolution of the STEM
image is not quite as good as that of the TEM image, but it could have
been improved by reducing the spot size. However, doing so would have
lowered the beam current, degrading the signal-to-noise ratio and increasing
image acquisition time. Nevertheless, the STEM image has higher resolution
and materials contrast than the SEM image. Additionally, it is not obscured
by the gold coating required in SEM sample analysis to prevent charging
artifacts.
 |
| Figure
6: (a) STEM, (b) TEM, and (c) SEM micrographs of a transistor. The
STEM image (at 30 kV) and the TEM image (at 200 kV) are of comparable
quality. |
Shallow-trench-isolation structures are shown in the 30-kV STEM and 200-kV
TEM micrographs in Figures 7a and 7b, respectively. The images show equal
levels of detail and contrast. Figure 8a presents a high-voltage (300
kV) STEM image captured using a HAADF detector, while Figure 8b presents
a low-voltage (30 kV) STEM dark-field image. The images show a planar
view of the same thin aluminum film. The lower-voltage image provides
better differentiation of the grain structure in the film than the higher-voltage
one.
 |
| Figure
7: (a) a 30-kV STEM micrograph and (b) a 200-kV TEM micrograph of
shallow-trench-isolation structures. The images show equal levels
of detail and contrast. |
 |
| Figure
8: (a) a high-voltage (300 kV) STEM image captured using a HAADF detector
and (b) a low-voltage (30 kV) STEM dark-field image. The images show
a planar view of the same thin aluminum film. |
Low-voltage STEM bright-field and dark-field images of polymer-resist
structures are illustrated in Figures 9a and 9b, respectively. Materials
with low atomic weight show poor contrast at higher beam energies. Since
the mean free path of an electron decreases with beam energy, lower beam
energies provide better mass thickness contrast, eliminating the need
to stain lighter materials.
Conclusion
STEM, together with automated sample preparation, promises to provide
significant analytical cost savings by offering high throughput, relatively
low capital expenditures, and personnel savings. Automation can cut average
sample preparation time by a factor of two—from more than two hours
to approximately one hour. Equally important, it may increase reliability
by nearly the same amount. In addition, because STEM imaging is faster
and easier than TEM imaging, the improvement in preparation throughput
can be realized as an improvement in overall analytical throughput.
 |
| Figure
9: Comparison between (a) low-voltage STEM bright-field and (b) dark-field
images of polymer-resist structures. |
A new state-of-the-art TEM can easily cost more than $2 million and must
be operated by top-level engineering personnel. A low-kV STEM, on the
other hand, can be added to an existing field-emission SEM for $50,000
and can be operated by technician-level operators. Even when the additional
cost of automated sample preparation is taken into consideration, the
STEM capital expenditure is less than half that of TEM. While these analytical
cost savings are significant in their own right, the true value of STEM
analysis lies in the impact of rapid analytical turnaround times on semiconductor
manufacturing operations. Fast recovery from process yield excursions
reduces yield loss, and fast development cycles for new products and processes
ultimately translate into greater market share and higher price margins.
Low-voltage STEM provides TEM-like resolution and material contrast at
a fraction of the cost of TEM. Automated sample preparation eliminates
much of the time and difficulty associated with conventional thin-sectioning
techniques and improves the overall reliability of the sample preparation
process. Together, low-voltage STEM and automated sample preparation offer
a viable, cost-effective solution for high-resolution imaging with the
capability to support semiconductor failure analysis well into the foreseeable
future.
Bryan
Tracy, PhD, has managed the materials characterization
laboratory at Spansion (Sunnyvale, CA) since 1996. The group performs
materials analysis to support the fab’s advanced process development
for flash memory devices. In 1984 he entered the semiconductor industry
and established TEM as a regular analytical technique at Intel’s
facility in Santa Clara, CA. In 1991, he joined AMD as section manager
in charge of SEM, TEM, and FIB instrumentation. Tracy holds 10 U.S. patents
and has published more than 25 technical papers. His interests focus on
the use of electron microscopy to characterize semiconductor materials.
In 1984 he received a PhD in material engineering from Rensselaer Polytechnic
Institute in Troy, NY. (Tracy can be reached at 408/749-4819 or
bryan.tracy@spansion.com.)
Kirstin
Alberi joined Spansion as a summer intern in 2003. She worked
with the materials characterization lab to investigate the application
of electron microscopy techniques in device failure analysis. She is a
graduate student in the materials science and engineering department at
the University of California, Berkeley, where she is researching thin-film
processing and nanostructure fabrication. In 2003 she received a BS in
materials science and engineering from the Massachusetts Institute of
Technology in Cambridge.
Shams
Tabrez is national sales and marketing manager for SELA USA (Sunnyvale,
CA). With more than 15 years of experience in semiconductor device manufacturing
and capital equipment, he has worked at Lam Research, National Semiconductor,
and other companies. Tabrez has published nearly 20 technical and commercial
papers and holds a U.S. patent. He has a BS in chemistry from the University
of Manchester (UK) and a master’s degree in chemical engineering
and economics from the Imperial College, University of London. He has
also performed executive MBA work at the University of Chicago Graduate
School of Business. (Tabrez can be reached at 408/736-3700 or shams@sela.com.)

MicroHome |
Search | Current Issue | MicroArchives
Buyers Guide | Media Kit
Questions/comments about MICRO Magazine? E-mail us at cheynman@gmail.com.
© 2007 Tom Cheyney
All rights reserved.
|