|
|
 |
Critical Material-Wafers
Monitoring
epitaxial resistivity profiles without wafer damage
Karen Woolford, Lee Newfield,
and Christopher Panczyk, SUMCO USA
A noncontact measurement
technique equals or betters the capabilities of current
metrologies, while eliminating the need to scrap wafer
samples.
Manufacturers
of advanced microprocessors, BiCMOS devices, and state-of-the-art
DRAMs look to epitaxial, or epi, wafers for the crystalline
silicon surfaces on which to build such devices. In response,
the wafer manufacturers must ensure that their products
meet customers' performance specifications and cost targets.
Pricing pressures have driven recent efforts to lower
the costs of producing epi wafers. One new technology
that promises to dramatically reduce costs is a metrology
tool capable of measuring epi-layer resistivity without
wafer contact, thereby eliminating the wafer scrap associated
with the traditional means of monitoring this specification.
Wafer
manufacturers have been using mercury-probe or capacitance-to-voltageSchottky
(HgCV) techniques to measure the resistivity of wafers
sampled from production runs. While these methods adequately
measure a wafer's resistivity profile (resistivity as
a function of depth into the silicon), any measured wafer
must be scrapped, because of surface damage or metallic
contamination, and manufacturers may lose as much as 4%
of a production run for such sampling. To be qualified
as a production-worthy alternative tool, however, the
noncontact system must be able to demonstrate equal or
improved measurement capability over a broad range of
devices compared with current metrology techniques, without
any wafer damage or contamination.
The
project described in this article evaluated the Epimet
Model 2 noncontact resistivity monitor from SemiTest (Billerica,
MA) for its suitability to monitor epi wafer production.
Designed for precision and accuracy, the system uses a
column of air in place of the mercury used in traditional
tools. Its failure modes are almost exclusively shutdown
or a zero reading, which is advantageous because there
is little or no measurement drift before failure. The
dual focus of the research was to determine the measurement
capability of this noncontact system and to answer the
key question of whether the technique alters the sampled
wafer in any way, causing it to be rejected.
Experimental
Considerations
Achieving
reliable measurements with the noncontact system requires
a surface preparation process analogous to what is required
for sample measurement on HgCV tools. While the surface
preparation technique for HgCV testing does not require
that product quality be maintained, however, wafers prepared
in the noncontact tool's surface pretreatment chamber
(PTC) must meet this criterion.
In
order to gauge the feasibility of the pretreatment and
measurement protocol, a two-part evaluation plan was implemented.
First, capability tests were conducted to determine the
measurement performance of the tool. Stability, bias linearity,
repeatability and reproducibility, and correlation to
the performance levels of existing HgCV tools were all
quantified and summarized.
Then
a battery of tests were performed to determine whether
any wafers measured by the noncontact resistivity tool
had been damaged. Evaluations included analyses of gate-oxide
integrity (GOI), measured by J*t testing (which measures
current density per time in coulomb/cm2); metal
contamination levels (MCL), measured by vapor-phase decomposition,
inductively coupled plasmamass spectrometry (VPD
ICP-MS); minority carrier lifetime, measured by microwave
photoconductive decay (µ-PCD); microroughness, measured
by scanning force microscopy (SFM); and particles added,
measured as light-point defects (LPDs) using a Surfscan
6200 from KLA-Tencor (San Jose). These tests cover potential
chemical, mechanical, and electrical alteration to the
silicon epi layer. The noncontact tool was designed to
measure 100200-mm wafers; 150-mm wafers were used
for the tests discussed here.
Measurement
Recipe Setup. When comparing the results from the
noncontact tool with those from HgCV testing, there are
several factors that need to be considered. Ideally, a
direct comparison can be obtained by measuring the same
points on the same wafer with both kinds of instrument.
Therefore the measurement pattern across the wafer for
the noncontact system was set up to be consistent with
the HgCV probe pattern, including the same edge exclusion.
Another factor that should remain constant is the method
used for averaging the dopant concentration before converting
results to resistivity data. For HgCV probes, an average
is taken of a section of the dopant concentration over
a specified depth region. Although the depth regions used
in the measurement recipes in the comparison testing were
not identical, they were representative of the flat profile
segment in the epi-layer dopant concentration over a fixed
depth.
Measurement
Capability Test Protocol. The protocol for measurement
capability testing was based on QS9000 guidelines for
a variable measurement system study. Samples represented
three resistivity ranges each for n- and p-type dopants,
as indicated in Table I. The QS9000 standard states that
measurement repeatability and reproducibility variation
must take up no more than 30% of the specification range,
with a target of 10%. This goal is challenging with HgCV
techniques because of the variability inherent in those
measurements.
Sample
ID |
Dopant
Type |
Epi-Layer
Thickness (µm) |
Resistivity
(W-cm) |
| A |
p |
4.50
|
1.77
|
| B |
p |
6.00
|
11.10
|
| C |
p |
17.00
|
28.50
|
| D |
n |
20.00
|
0.96
|
| E |
n |
19.00
|
5.48
|
| F |
n |
12.00
|
12.60
|
|
| Table
I: Epitaxial silicon layer information by sample ID.
|
Product
Integrity Verification. Because HgCV test wafers are
scrapped, no correlation testing was required in the area
of postmeasurement product quality. However, since the
goal of implementing noncontact monitoring is to measure
actual product wafers, it was important to determine whether
the sampled wafers are damaged by any step in the measurement
protocol. In this study, the LPD, µ-PCD, and MCL
analyses were performed by partitioning the wafer process
flow so that deviations in wafer quality could be traced
to individual process steps. A GOI test also was performed
to verify that no other problems existed after wafer resistivity
was measured by the noncontact tool.
Measurement
Capability Test Results
Stability.
The standard deviation method was used to determine the
noncontact tool's stability when measuring n- and p-type
dopants in various resistivity ranges. Table I lists the
sample parameters. Three measurements were taken on each
sample every 12 hours over a 25-day period. The QS9000
acceptance criterion is that the measurement-process ratio
is <30% of the product specification; and the noncontact
resistivity tool passed for all six samples. The results
are summarized in Table II, with typical data plots shown
in Figures 1 and 2.
|
Sample
ID
|
s
Measurement/
s process (%)
|
|
A
|
20.3
|
|
B
|
9.23
|
|
C
|
13.6
|
|
D
|
15.5
|
|
E
|
14.5
|
|
F
|
19.1
|
|
|
Table II: Stability
results by sample. The acceptance criterion is that
the ratio of variabilities (measurement to process)
must be less than 30%.
|
| |
|
|
| Figure
1: Stability results for sample A: (a) x-bar data,
and (b) range data. |
| |
 |
| Figure
2: Stability results for sample B: (a) x-bar data,
and (b) range data. |
Bias
Linearity. Bias is the difference, or offset,
between a measurement and a known value, while linearity
is the fit of a regression line to different offsets over
the tolerance of the tool. Ideally, the bias should be
the same whether one measures resistivity at 1 or 60 W-cm.
The QS9000 standard mandates that the regression line
slope should be no greater than 0.03 and the fit of measured
to known values should be greater than 0.95. Bias linearity
is normally evaluated using an independently established
standard (from the National Institute of Standards and
Technology [NIST]); in this study, three p-type substrate
standards from VLSI Standards (San Jose) were usedRS801,
RS810, and RS 860which have resistivity values of
0.861, 10.48, and 56.40 W-cm,
respectively. The regression line was calculated and compared
with these standard values, yielding QS9000-compliant
results, as seen in Figure 3.
|
|
| Figure
3: Bias linearity data for the noncontact resistivity
tool measured against p-type VLSI standards of 0.861,
10.48, and 56.04 W-cm.
|
Repeatability
and Reproducibility. The repeatability and reproducibility
(R&R) of the tool measurements were assessed using
the average-and-range method, which is often used to determine
operator-to-operator variation. Because the noncontact
system features automatic loading, the component of variation
usually assigned to the operators instead represents a
factor of drift or sensitivity to evolving environmental
conditions (humidity, barometric pressure, and so forth).
All six samples were assessed against typical customer
specifications for epi wafers. The acceptance criterion
is that the ratio of R&R to the specification tolerance
must be less than 30%. The noncontact tool showed acceptable
repeatability and reproducibility in all cases. These
results are summarized in Table III.
| Sample
ID |
R&R
Spec.
Tolerance (%)
|
| A |
16.80 |
| B |
7.28 |
| C |
14.20 |
| D |
18.40 |
| E |
13.70 |
| F |
7.50 |
|
| Table
III: R&R results by sample. The acceptance criterion
is that the ratio of R&R to the specification
tolerance must be less than 30%. |
Correlation.
To assure that epi wafers manufactured to customer specifications
are not affected by a change in metrology, it is critical
to correlate the capabilities of the new instrument to
those of existing instruments. For this purpose, measurements
were taken on five sample wafers covering the full resistivity
range over five days using both the noncontact tool's
protocol and the HgCV protocol. In this comparison, a
single-point calibration curve was established using the
results for a high-resistivity wafer measured with the
HgCV method as the standard or baseline.
Considering
the large variability typically experienced in HgCV reading,
this clearly is not the best means of calibrating a more-precise
gauge. However, it is reasonable for purposes of comparing
the two techniques. (Under production conditions, the
noncontact tool is calibrated to NIST standards.) When
the daily measurements were averaged and correlated, the
resultant linear fit yielded a slope of 1.0155, with R2
= 0.9958, as shown in Figure 4. The variation in the HgCV
data (shown horizontally) was much broader than the variation
in the noncontact tool data (shown vertically). This correlation
meets SUMCO's criterion for technology transfer.
|
|
| Figure
4: Correlation between noncontact-tool and HgCV measurements
for n- and p-type samples. Correlation between the
two gauges met in-house criterion for product and
recipe transfer. |
Product
Integrity Verification Results
Samples
for each product integrity test were taken from one run
of material at four different points: directly after epitaxial
deposition; after PTC processing; after PTC processing
and measurement; and after PTC processing, measurement,
and SC-1/SC-2 cleaning. The epi-only samples served as
reference wafers.
Particle
Performance. Light-point defects are a measure of
the number and size of particles on a wafer surface. Figure
5 shows the average number of LPDs in six size ranges
from 0.12 to 20 µm on the reference wafer and on
samples subjected to various processes. LPD specifications
vary greatly between customers; however, the data indicate
that wafers conditioned with PTC or both pretreated and
measured on the noncontact tool would be acceptable for
most specifications. All samples would pass current international
particle specifications for 150-mm wafers.
|
|
| Figure
5: LPD test results for a reference wafer and samples
from three process-flow points. The data for bins
1 through 6 represent increasingly large particles,
with bin 1 being ≤0.12 µm and bin 6 ≤20
µm. |
Carrier
Lifetime. To determine if one or a combination of
the process steps in the measurement protocol would cause
carrier lifetime degradation, two wafers from each of
the four sampling points were evaluated using µ-PCD.
The data shown in Figure 6 suggest that none of the partitioned
process steps are associated with minority carrier lifetime
degradation. In fact, wafers subjected to some or all
of the protocol exhibited better lifetimes (on average)
than did the reference wafers and cleaned samples.
|
|
| Figure
6: Results of µ-PCD measurements on two reference
wafers and two wafers each from three sampling points.
All samples showed acceptable minority carrier lifetimes.
|
Metal
Contamination. VPD ICP-MS testing was performed to
evaluate the surface metal contamination contributed by
each step in the measurement protocol. As shown in Figure
7, measured surface concentrations for all elements were
below 1.5 x
1010 atoms/cm2. While specifications
vary by customer and element, this threshold is considered
acceptable by the industry.
|
|
|
Figure 7: VPD ICP-MS
results for a reference wafer and samples from three
process-flow points. Specifications vary by element;
however, all of the metal contamination levels shown
would be considered acceptable.
|
Electrical
Effects and Surface Morphology. Effects of the noncontact
tool's measurement protocol on gate-oxide integrity and
surface roughness also were determined, using J*t measurements
and SFM. The resulting data, shown in Figures 8 and 9,
indicate that there is no difference between measured
epi wafers and nonmeasured wafers.
|
|
|
Figure 8: Typical
J*t yield-map data obtained from GOI testing. All
pretreated, measured, and cleaned wafers tested
showed acceptable yields, comparable to those of
reference samples.
|
| |
 |
| Figure
9: RMS roughness data measured using SFM on a 1 x
1-µm area on reference wafers and wafers measured
using the noncontact tool's protocol. |
Conclusion
The
capability test results reported here indicate that resistivity
measurements performed by the noncontact system are more
repeatable and stable than those performed using the current
HgCV protocol. The tool exhibited less variation in all
measurement ranges explored in the study, which represent
typical MOS epi wafer product specifications for n- and
p-type dopants. The system also exceeded the performance
and capability of the HgCV technique for resistivity measurements
above 1.7 W-cm.
In
addition, product integrity verification testing demonstrated
that use of the tool was not associated with any contamination
problems, and a gate-oxide integrity test indicated that
it had no deleterious effect on device manufacturing yield.
It should be noted, however, that product integrity is
closely tied to the tool's maintenance history. Poor handling
practices or incorrect operation can lead to hardware
contamination, resulting in deviations from the performance
levels captured in this study. Extreme care and rigorous
requalification of the system is needed when maintenance
activities breach the pristine wafer-handling areas of
the PTC or the measurement chuck.
When
the new metrology is implemented in ongoing, continuous
manufacturing operations, reliability will become as great
a concern as the system's capabilities and the maintenance
of product quality. The tool's overall reliability and
uptime measured by mean time to failure and mean time
to repair will be assessed in the next phase of the work.
This
article is based on a poster paper originally presented
at the 13th Annual Advanced Semiconductor Manufacturing
Conference (ASMC) on April 30, 2002, in Boston. The authors
would like to thank Greg Martel, a technical representative
from SemiTest who contributed his expertise on the Epimet
noncontact resistivity monitor.
Karen
Woolford is a quality engineer at SUMCO USA in Fremont,
CA, where she is calibration coordinator and oversees
the quality technicians. She focuses on optimizing the
performance of the various metrology tools used to monitor
the epitaxial manufacturing process. Over the past eight
years at the company, she has expanded her epitaxial wafer
processing background in manufacturing, R&D, and quality.
She has four years of direct work experience with the
Epimet technology, including in the area of manufacturing
integration. Woolford has been a member of the American
Society for Quality since 2001. She received a BA in physical
sciences, opt B, from the University of California at
Berkeley in 1994. (Woolford can be reached at 510/440-3385
or karen. woolford@sumcousa.com.)
Lee
Newfield is quality engineering manager of SUMCO USA
in Fremont, CA. He is in charge of metrology, statistical
process control, and customer reporting. Newfield has
23 years of experience in silicon wafer engineering, including
crystal growth through epitaxial deposition. He is a member
of the American Society for Quality, from which he received
certifications for quality engineering, quality auditing,
and quality management. (Newfield can be reached at 510/440-3325
or newfield@sumcousa.com.)
Christopher
Panczyk, PhD, is the director of quality at SUMCO,
where he is responsible for the management of quality
engineering and quality assurance activities and programs
for multiple manufacturing facilities on the West Coast.
He is also the acting director of operations for the Fremont
division, managing both the manufacturing and maintenance
departments. Panczyk started his career in R&D at
Sumitomo Sitix Silicon in 1996. He is a member of the
Materials Research Society. He received a BS in chemical
engineering from the University of Illinois at Urbana-Champaign
and an MS and a PhD in chemical engineering from Purdue
University in West Lafayette, IN. (Panczyk can be reached
at 510/440-3334 or cpanczyk@sumcousa.com.)

MicroHome |
Search | Current Issue | MicroArchives
Buyers Guide | Media Kit
Questions/comments about MICRO Magazine? E-mail us at cheynman@gmail.com.
© 2007 Tom Cheyney
All rights reserved.
|
|
 |
 |
 |
|