Analyte
|
Unspiked
Sample 1 (ppt) |
Unspiked
Sample 2 (ppt |
Average
Unspiked |
Spiked
Sample 1 (ppt) |
RSD
(%) |
Spiked
Sample 2 (ppt) |
RSD
(%) |
%
Recovery Spiked Sample 1 |
%
Recovery Spiked Sample 2 |
% Recovery
Range |
MDL |
|
Aluminum
|
2.1
|
3.2
|
2.6
|
7.0
|
9
|
7.2
|
11
|
86
|
91
|
5
|
5.9
|
|
Antimony
|
0.41
|
0.39
|
0.40
|
4.7
|
16
|
4.7
|
7
|
86
|
87
|
1
|
0.55
|
|
Arsenic
|
2.2
|
2.4
|
2.3
|
6.4
|
15
|
6.4
|
13
|
83
|
82
|
1
|
1.6
|
|
Barium
|
0.34
|
0.32
|
0.33
|
4.9
|
8
|
4.7
|
4
|
91
|
87
|
4
|
0.45
|
|
Boron
|
12.2
|
11.9
|
12.0
|
16.7
|
14
|
17.6
|
14
|
92
|
111
|
19
|
7.0
|
|
Calcium
|
2.5
|
2.8
|
2.6
|
6.6
|
7
|
6.9
|
14
|
80
|
86
|
6
|
1.4
|
|
Chromium
|
0.86
|
1.0
|
0.9
|
5.2
|
9
|
5.4
|
7
|
85
|
89
|
4
|
0.70
|
|
Cobalt
|
0.26
|
0.35
|
0.3
|
5.0
|
5
|
4.9
|
4
|
93
|
92
|
1
|
0.56
|
|
Copper
|
2.3
|
2.9
|
2.6
|
7.0
|
20
|
7.1
|
12
|
88
|
91
|
3
|
2.4
|
|
Iron
|
9.2
|
9.8
|
9.5
|
14.7
|
6
|
14.2
|
3
|
103
|
94
|
9
|
2.7
|
|
Lead
|
1.2
|
1.2
|
1.2
|
6.0
|
8
|
5.9
|
4
|
96
|
94
|
2
|
0.69
|
|
Lithium
|
0.47
|
0.44
|
0.5
|
5.7
|
9
|
5.0
|
12
|
104
|
91
|
13
|
1.2
|
|
Magnesium
|
0.66
|
0.87
|
0.8
|
5.6
|
5
|
5.5
|
16
|
96
|
94
|
2
|
1.1
|
|
Manganese
|
0.95
|
0.81
|
0.9
|
4.9
|
8
|
4.8
|
7
|
80
|
78
|
2
|
1.6
|
|
Nickel
|
3.0
|
3.0
|
3.0
|
7.1
|
12
|
7.4
|
15
|
83
|
89
|
6
|
1.5
|
|
Potassium
|
5.1
|
5.2
|
5.1
|
9.1
|
12
|
9.2
|
7
|
79
|
81
|
2
|
1.3
|
|
Sodium
|
4.5
|
4.5
|
4.5
|
9.6
|
3
|
9.9
|
7
|
103
|
109
|
6
|
1.8
|
|
Tin
|
5.7
|
6.1
|
5.9
|
10.8
|
12
|
10.9
|
8
|
97
|
100
|
3
|
2.4
|
|
Titanium
|
4.7
|
5.2
|
5.0
|
10.1
|
10
|
10.3
|
10
|
103
|
107
|
4
|
2.3
|
|
Vanadium
|
0.23
|
0.29
|
0.3
|
4.8
|
7
|
4.8
|
15
|
90
|
91
|
1
|
0.32
|
|
Zinc
|
5.8
|
6.5
|
6.1
|
11.3
|
11
|
11.5
|
9
|
103
|
108
|
5
|
3.0
|
| Table VII: Summary of spike recovery and MDL data for 31% ultra-high-purity hydrogen peroxide analyzed using DRC ICP-MS technology. |